Add to compare : PUE C315 Indicator

Selected language English English
Change language
Change language page on Deutsch DeutschChange language page on Español EspañolChange language page on Français FrançaisChange language page on Italiano ItalianoChange language page on Polski PolskiChange language page on Türkiye Türkiye ČeskyChange website to USA USAChange language page on 中文 中文

PUE C315 Indicator

PUE C315 Indicator - Functions of an indicator PUE C315 series: measuring units: [g], [kg], [N], [ct], [lb]; tarring in whole measuring range; automatic tare, tare memory; parts counting with equal part mass; +/- control of mass with set standard; percent setup with set standard; averaging of weighing result, digital filter; control of power supply from batteries; scale timer switch off; adjustable backlight on powering by batteries; adjustable baud rate between 1200-38400 bit/s; continuous data transmission through RS 232; manual or automatic mode for RS 232; weighing loads with switched off autozero function; measurement of maximum force or maximum mass placed on weighing pan; measurement of force influencing the weighing pan (in Newtons); start mass control; slot for connecting an additional LCD display; totalizing mode.  The indicator PUE C315 series enables designing printouts according to individual needs
This image is for reference only.

Product code: WX-008-0150
Request for quotation

Overview

Functions of an

indicator

PUE 
C315 series:
  • measuring units: [g], [kg], [N], [ct], [lb];
  • tarring in whole measuring range;
  • automatic tare, tare memory;
  • parts counting with equal part mass;
  • +/- control of

    mass

    with set standard;
  • percent setup with set standard;
  • averaging of

    weighing

    result, digital filter;
  • control of power supply from batteries;
  • scale timer switch off;
  • adjustable backlight on powering by batteries;
  • adjustable baud rate between 1200-38400 bit/s;
  • continuous data transmission through RS 232;
  • manual or automatic mode for RS 232;
  • weighing

    loads with switched off autozero function;
  • measurement of maximum force or maximum

    mass

    placed on weighing pan;
  • measurement of force influencing the weighing pan (in Newtons);
  • start mass control;
  • slot for connecting an additional LCD display;
  • totalizing mode.

 The

indicator

PUE C315 series enables designing printouts according to individual needs.

Functions

Plus/Minus Control

Plus/Minus Control

Control of sample's mass carried out with reference to specified MIN and MAX thresholds. The weight is considered to be correct if it is comprised within threshold values. A graphic visualisation of weighing result is displayed in a form of a bar graph.
Percent Weighing

Percent Weighing

Percent Weighing function is used for comparision of measured products with mass standard. Mass of a mass standard may be a numeric value taken from a database or it may be determined through a measurement process. Each measured product is compared to mass standard, mass of which is presumed as a model 100% ideal mass. For products weighing less than the mass standard, obtained results are lower than 100%, for products weighing more, the obtained results are greatly  exceeded.
Totalizing

Totalizing

Function adding new mass value to already obtained one. This allows to determine mass content of the mixture, being a composition of number of different consistence components. Totalizing is limited by maximum capacity parameter.
Parts counting

Parts counting

Function using mass measurement for determination of  measured items quantity. Mass of a single item is required for this process. It may be either estimated through weighment or taken from a database. For items counting the following algorithm is used: all items mass / single item mass = quantity. Function operation is supperted by a mechanism of Automatic Correction of Accuracy. This allows to update single item mass in course of the process. To a certain extend Automatic Correction of Accuracy eliminates error which may be a result of different mass values of seemingly alike single elements. For industry solutions items counting may be simultaneously carried out with checkweighing and dosing thus industry solutions feature audio signalling base informing that specified number of items has been weighed. It is possible to apply weighing systems using few platfroms of different MAX capacities and different accuracies.
Internal battery

Internal battery

Internal and rechargeable battery of SLA 6 V type (sealed lead acid) enables device operation without connecting the device to the mains (9h operation). The weighing instrument controls battery status and displays a respective pictogram on the screen.
Peak hold

Peak hold

Function registering the greatest temporary indication occurring in course of the weighing process. It is most often used in medical scales in order to hold the measurement indication.
GLP Procedures

GLP Procedures

Diagnostic function allowing to objectively document performed measurements. GLP procedures may be either presented in a short report form or extended one.
Animal weighing

Animal weighing

Process of mass determination for a product which may unwillingly reposition within the weighing pan. Mass determination in such cases requires much longer period of time when compared to typical weighing process. It is the user who defines period of time needed for control of measured product mass. The user can thus optimize the function depending on the measured product characteristics.
ALIBI Memory

ALIBI Memory

"ALIBI" memory enables saving and storing up to 100000 measurements.  Measurements are saved automatically upon each pressing of the key. There is no need for any additional operations or changing settings.
The following data related with the measurements is saved: measurement date; measurement time; result (mass); tare value; operator carrying out the measurement (logged in); product name.
The software saves the measurement in a so called loop, i.e. when the measurement no. 100001 is saved, the measurement 1 gets automatically deleted from the memory.
Measurements saved to balance memory cannot be deleted.
You can preview and print out the data saved to the ALIBI database.

Datasheet

Maximum quantity of verification units

6000 e

 

OIML Class

III or IV

 

Display

LCD (backlit)

 

Protection class

IP 43

 

RS232 Interface

1× standard, 1× option

 

Wi-Fi® Interface

1× standard

 

Power supply

100 ÷ 240 V AC 50 / 60 Hz

 

Power supply

detachable adapter, akumulatory NiMH 6×AA

 

Optional power supply

12V DC and battery

 

Operation time on batteries

max 7h

 

Maximum quantity of divisions from converter

838 860 ×10

 

Maximum increase of signal

19.2 mV

 

Minimum volatge per verification unit

0.4 µV

 

Maximum voltage per verification unit

3.25 µV

 

Minimum load cell impedance

50 Ω

 

Maximum load cell impedance

1200 Ω

 

Load cell excitation voltage

5V DC

 

Connection of load cells

4 or 6 wires + shield

 

Operating temperature

-10 ÷ +40

 °C

Storage temperature

-10 ÷ +50

 °C

Atmospheric humidity

10% ÷ 80% RH no condensation

 

Keypad

microswitch

 

Number of buttons

5

 

Housing

ABS plastic

 

Device dimensions

181×136×60

 mm

Packaging dimensions

220×190×90

 mm

Net weight

0.6

 kg

Gross weight

1.2

 kg

Multiple range

1 or 2 ranges

 

Additional LCD display

WD-4/1 (option)

 

Documentation

Technical specification
Publications and teaching materials
User Manual
(0.84 MB)
Save See the publication
Manuals
(0.30 MB)
Save See the publication
Data Sheet
(1.00 MB)
Save See the publication
Conformity Declaration
(0.04 MB)
Save See the publication
C315 Platform Scales
(2.13 MB)
Save See the publication

Compatible with

Waterproof stainless steel platforms
Mild steel powder coated weighing platforms
Mild steel powder coated weighing platforms
Scales Editor 2.0
RW Tool
R-LAB
RAD-KEY
RADWAG Connect
RADWAG Development Studio
LabVIEW Driver
Receipt printer
RS 232 cables (scale - EPSON printer)
RS 232, RS 485 cables
Cigarette lighter receptacle power supply cables
Stands, wall mounting kits and mounting brackets
RS232 1345 Converter
RS232 1339 Converter
Power loop output
Displays

Radwag Balances and Scales

Torunska 5
26-600 Radom
Poland
tel.: +48 48 386 60 00
e-mail: export@radwag.com
TIN: PL7960000327
Vimeo Radwag YouTube Radwag Facebook Radwag
© 2019 RADWAG Balances and Scales. All rights reserved.

Watch more
RADWAG videos

movie