AP-12.5Y Automatic Device for Multichannel Pipette Calibration
- two calibrated sensors: THB S and THB W,
- two weighing pans: standard, and for balance test and calibration,
- two pumps: a single-channel pump and a four-channel pump
AP-12.5Y Automatic Device for Multichannel Pipette Calibration
⤢
ø26 + automat⤢
ø26 + automatMaximum capacity [Max] | 52 g |
Readability [d] | 0,01 mg |
Tare range | -52 g |
Standard repeatability [5% Max] | 5 µg |
Standard repeatability [Max] | 10 µg |
Standard minimum weight (USP) | 10 mg |
Standard minimum weight (U=1%, k=2) | 1 mg |
Linearity | ±0,03 mg |
Stabilization time | 4 s |
Adjustment | internal (automatic)
|
Typical measurement time for a 12-channel pipette with constant volume, according to ISO 8655 (for d=10 µg)* |
under 20 minutes
|
Typical measurement time for a 12-channel pipette with variable volume, according to ISO 8655 (for d=10µg)* |
under 60 minutes
|
Maximum volume of dishes | 12 ml |
Display | 10″ graphic colour touchscreen
|
Weighing pan dimensions | ø26 + automat
|
Packaging dimensions | 605×560×775 mm |
Net weight | 17 kg |
Gross weight | 30 kg |
Communication interface | 2×USB-A, USB-C, HDMI, Ethernet, Wi-Fi®, Hotspot
|
Operating temperature | +10 – +40 °C |
Operating temperature change rate | ±0,3 °C / 1 h (±1 °C / 8 h)
|
Relative humidity | 20% – 80%
|
Relative humidity change rate | ±1% / h (±4% / 8 h)
|